Artigo

Determination of mercury in liver of marine tetrapods by cold vapor atomic fluorescence spectrometry and inductively coupled plasma mass spectrometry in biological samples: A systematic comparison between the two techniques

Resumo

Abstract:

This study compares the techniques cold vapour atomic fluorescence spectrometry (CV-AFS) and inductively coupled plasma mass spectrometry (ICP-MS) for the determination of Hg in biological samples (hepatic tissue from marine tetrapods). Acid digestions were performed for the samples and aliquots of them were analysed by ICP-MS and CV-AFS. To minimize the Hg memory effect in its determination by ICP-MS, gold (200 μg L-1 in 5% v v-1 HCl) was added to the sample digests. The detection limits were 0.02 mg kg-1 for ICP-MS and 0.01 mg kg-1 for CV-AFS. Accuracy was assessed by the calculation of the recovery obtained from TORT-3 (lobster hepatopancreas) certified reference material analysis. Accurate results were obtained for both CV-AFS and ICP-MS. The comparison of the results obtained for hepatic tissue with and without lyophilization showed a significant difference among them (at 95% of confidence level), evidencing losses of Hg in the lyophilization process. © 2021 Sociedade Brasileira de Quimica. All rights reserved.

Descrição

Palavras-chave

Citação

ISSN

Coleções

Avaliação

Revisão

Suplementado Por

Referenciado Por

Licença Creative Commons

Exceto quando indicado de outra forma, a licença deste item é descrita como Attribution-NonCommercial-NoDerivs 3.0 Brazil